Sound velocity measurement and acoustic levitation experiment

FOB Reference Price:Get Latest Price
$500.00 - $2,000.00 / Set | 1 Set/Sets (Min. Order)
Shipping:
Support Sea freight
Report Suspicious Activity
Overview
Quick Details
Subject:
Physics
Brand Name:
Brolight
Model Number:
SEK-8107
Place of Origin:
Zhejiang, China
Supply Ability
Supply Ability:
100 Unit/Units per Month
Packaging & Delivery
Packaging Details
Carton
Port
Shanghai Airport or Ningbo Port
Lead Time :
15 - 30 days

 

Sound velocity measurement and acoustic levitation experiment

  SEK-8107

 

 

 

DETAILS

The instrument is composed by acoustic levitation experimental device and signal source. It can operate on acoustic levitation experiment and also can repeat traditional measuring velocity of sound experimental context

Experimental contents

1. Master resonance interference method, phase-comparison method and time-difference method to measure the velocity of sound in different medium.

2. Master using acoustic levitation phenomenon to measure the velocity of sound.

Features

a) Use ultrasonic to suspend the object in the air medium.

b) Observing the resonance and sine distribution in the air medium can be a demonstrative experiment for acoustic levitation. And also it can be used for measuring ultrasonic wavelength by acoustic levitation and the experimental context by repeating traditional measuring the velocity of sound experiment.

Technical parameters

Superposed layer

Levels:1-6 level

Length between levels: 4.5mm

Maximum of suspended distance:25-28mm

Tester

Piezoelectric ceramic transducer resonance frequency:35±3kHz.

Tolerable continuous electric power must bigger than 15W.

Power signal source

Range of frequency:25kHz~50kHz.

Display of frequency: LED five-digit display.

Pulse modulation signal source

Frequency:38.5kHz, Pulse width:27μs,Pulse period:60ms.

Count-up timer

Range of count timer):1μs~1s; Resolution:1μs.

Measure method

Stationary wave method, phase method, time-difference method and acoustic levitation method.

Measuring medium

Air, solidity need to collocate by user.

 

Typical experimental context and data

Record and calculate the heating conductivity coefficient of different materials.

a) Measuring the velocity of sound wave in the air by stationary wave method.

                                Data figure

 

 Input voltage:15V

Input frequency:37.855kHz

Environment temperature:20

Measure times

1-4

5-8

9-12

Reading on rod when standing wave appears(mm)

25.26

43.55

61.85

29.84

48.13

66.43

34.41

52.70

71.01

38.98

57.27

75.59

 

Instrument: Sound velocity measuring combination instrument Unit 1 set,  dedicated signal source Unit 1 set,  dual trace oscilloscope  Unit 1 set.

Average  λ =9.156mm,Experimental value:Ve = f x λ = 37.855 x 103 x 9.156 x 10-3 = 346.60m/s.

Relative error: E=|Ve-Vt|/Vt x 100% = |346.60-343.37|/343.37 x 100% = 0.94%

b) Measuring the velocity of sound wave in the air by phase method.

Data figure

 

 

Instrument:  Sound velocity measuring combination instrument Unit 1 set, dedicated signal source Unit 1 set,  dual trace oscilloscope  Unit 1 set.

Averageλ=9.154mm,Experimental value:Ve = f x λ = 37.910 x 103 x 9.154 x 10-3 = 347.03m/s .

Relative error: E=|Ve-Vt|/Vt x 100% = |347.03 - 344.54|/344.54 x 100% = 0.72% 